PTFE Membranes

PTFE Membranes

PTFE Laminated Membrane Filters

PTFE Laminated Membrane Filter Applications

Filtration of strong acids and aggressive solutions

Venting applications

Phase separations

Aerosol samplings

Filtration of aqueous solutions (when prewetted with methanol)

Laminated PTFE Membrane Filter Specifications

Performance by Pore Size

Pore Size Rating (µm)Thickness (µm)Air Permeability1
(L/min·cm² at 70mbar)
Water Entry Pressure
(psi; bar)
IPA Bubble Point
(psi; bar)
0.1-0.2152-2540.12-0.43>60; 4.14≥17; 1.17
0.2076-1520.26-0.55>37; 2.55≥17; 1.17
0.4564-1270.60-1.19>11; 0.76≥10; 0.69
1.076-1272.56-5.97>4; 0.28>8; 0.55

1 Values calculated assuming linear relationship between flow and differential pressure

* NOTE: 

QP (Polyester) products max operating temp is 170°C for short term use and 130 – 150°C for continuous use.

QL (Polypropylene) products max operating temp is 130°C for short term use and 80°C for continuous use.

PTFE products max operating temp is 260°C.

PTFE Unlaminated Membrane Filters

Unlaminated PTFE Membrane Filter Specifications

Performance by Pore Size

Pore Size Rating (µm)Thickness (µm)Air Permeability1
(L/min·cm² at 70mbar)
Clean Water Flow2
(mL/min·cm² at 0.7bar)
Water Entry Pressure (psi)Alcohol Bubble Point3(psi)
0.225-510.3-0.915NT19-26
0.45 25-510.5-1.4NT>4513-23
1.0 203-3050.9-1.161-925.5-6.51.0-1.4
3.075NT300NT1.9
5.0 152-2541.1-1.392-1533.5-4.50.8-1.2
10 130 >1.2>94NT0.7 

Notes:

Airflow values calculated assuming linear relationship between volumetric airflow and differential pressure.

Water flows measured after prewetting with alcohol and calculated assuming linear relationship between flow and differential pressure.

Bubble points for 0.2µm, 0.45µm, and 3.0µm pore sizes measured using IPA; all others measured using ethanol.

Aspire Laminated ePTFE Membranes

Applications:

Filtration of strong acids and aggressive solutions

Venting applications

Phase separations

Aerosol samplings

Membrane Distillation

Laminated PTFE filters can also be used to filter aqueous solutions when prewetted with methanol.

Aspire Laminated ePTFE membrane Specifications*

Product CodeBacking LayerPore Size (µm)Thickness in Mil (µm)Air Permeability
(ft3/min·ft2 @ 125 Pa)
Water Entry Pressure (psi)IPA Bubble Point (psi)Surface Treatment
QL217Polypropylene0.1-0.26-10 (152-254)0.07-0.25>60>17Hydrophobic
QP952Polyester0.25-9 (127-229)0.15-0.35>50≥17Hydrophobic
QP955Polyester0.25-9 (127-229)0.05-0.15>60>20Hydrophobic
QL231Polypropylene0.26-10 (152-254)0.12-0.30>45>17Hydrophobic
QL822Polypropylene0.455-9 (127-229)0.10-0.45>37>15Hydrophobic
QL211Polypropylene0.458-12 (203-305)0.6-1.2>21>10Hydrophobic
QL827Polypropylene0.454-8 (102-203)0.30-0.65>12N/AOleophobic
QP909Polyester1.06-10 (152-254)0.3-0.8>12>6Hydrophobic
QL209Polypropylene1.06-10 (152-254)0.6-1.2>18>5Hydrophobic
QL208Polypropylene3.06-10 (152-254)1.5-3.5>5>1Hydrophobic
QL207Polypropylene5.06-10 (152-254)3.0-5.5>5>0.5Hydrophobic
QL230Polypropylene5.06-10 (152-254)4.8-8.0>1>0.5Hydrophobic
QP930Polyester5.03-5 (76-127)3.5-7.5>2>0.6Hydrophobic

*In April 2018, several specifications for individual Aspire PTFE Membranes were revised based on improved characterization equipment and procedures. Membrane characterization has no impact on the manufacturing process for the membranes, and no change in performance should be expected.

* NOTE: 

QP (Polyester) products max operating temp is 170°C for short term use and 130 – 150°C for continuous use.

QL (Polypropylene) products max operating temp is 130°C for short term use and 80°C for continuous use.

PTFE products max operating temp is 260°C.

Advantec Unlaminated PTFE Hydrophilic Membranes

Summary of Features:

Maximum chemical and pH resistance

High flow rates with minimal aqueous extractables (<0.3wt%)

Optically clear when wet with water

Unlaminated 

Ideal for HPLC and other mixtures of aqueous and organic solvents

Applications:

Chemicals, Petrochemicals, and Fertilizers

Computers, Semiconductors, and Electronics

Food and Beverage

Glass and Glass Products

Healthcare and Hospitals

Metals and Metals Fabrication, Mining (Coal, Minerals, Metals)

Pharmaceutical and Medicine

Power Generation

HPLC Seperations

Advantec General Hydrophilic PTFE Filter Specifications Product Performance:

Pore Size (μm)0.100.200.501.00
Bubble Point (MPa)ᵇ≥0.38 (≥55.1 psi)≥0.24 (≥34.8 psi)≥0.14 (≥20.3 psi)≥0.083 (≥12.0 psi)
Flow Rate (H2O)¹14213973
Flow Rate (Air)²1.62.12.95.7
Porosity (%)ᶜ71717983
Thickness (µm)35353535
Max. Operating T° (°C)100100100100

Notes:

Flow Rate indicates initial flow rate at 10 psi using a KGS 47 filter holder:</span ></em >

</strong >Water:</span > (mL/min/cm2) using prefiltered (0.1μm) water </span ></em >

Air: (L/min/cm2</sup >) using prefiltered nitrogen</em >

Bubble Point is the minimum pressure required to force air through a membrane which has been pre-wet with water.</span ></em >

Porosity refers to the percent open area.</span ></em >